Pipette Module Specification

Technical Data

Pipette Module Target Volume (μL) Systematic Error (%) Random Error (%)
Single Channel 50μL 1* 5 5
10 1.5 0.8
50 1 0.4
Single Channel 250μL 5 3 2.5
50 1.4 0.6
200 1.5 0.4
Single Channel 1000μL 100 1.0 0.3
200 0.8 0.3
1000 0.8 0.3

* dispensed into liquid.

Supported Tips

Manufacturer Product Code Description
Thermo Fisher 901-261 20µl, Clear, Sterile
902-021 50µl, Clear, Sterile, Filtered
902-261 50µl, Clear, Sterile
903-021 200µl, Clear, Sterile, Filtered
903-261 200µl, Clear, Sterile
904-021 1000µl, Clear, Sterile, Filtered
904-261 1000µl, Clear, Sterile
Corning Axygen TTF-20-C-HTR-S 20µl, Clear, Sterile
TT-50-C-HTR 50µl, Clear, Sterile
TTF-200-C-HTR-S 175µl, Clear, Sterile, Filtered
TT-200-C-HTR-S 200µl, Clear, Sterile
TTF-1000-C-HTR-S 1000µl, Clear, Sterile, Filtered
TT-1000-C-HTR-S 1000µl, Clear, Sterile
Starlabs R1181-5810 50µl, Clear, Filtered
R1128-5700 50µl, Clear
R1128-0810 200µl, Clear, Filtered
R1181-0700 200µl, Clear
R1128-2810 1000µl, Clear, Filtered
R1181-2700 1000µl, Clear

Note: The systematic and random error values can be achieved only under strictly controlled conditions during test as per ISO 23783 Annex D. The best compatibility is achieved when combining i.prep pipette modules with supported tips. The systematic error and random error verification results, in tests, have been achieved using configurations available in the latest  and i.prep firmware release. Due to continuous product development the systemic and random error values are subject to change without notice.